*** MIDAS logfile (05SEPpl1.2) *** Mo, 26 Dec 2005 14:41:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 26 Dec 2005 14:42:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 051226 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.48 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_051226 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1691.4 1691.8 830.1 0.11 0.12 2.04 2 6.16 3377.0 3376.7 1622.0 -0.13 -0.14 2.08 3 12.31 6755.9 6755.6 3223.5 -0.13 -0.14 2.10 4 18.47 10136.0 10135.3 4888.0 -0.12 -0.13 2.07 5 27.70 15208.3 15209.8 7454.0 -0.09 -0.09 2.04 6 36.93 20290.3 20290.2 9855.3 -0.04 -0.04 2.06 7 46.16 25372.9 25372.6 12198.9 0.00 -0.00 2.08 8 61.54 33866.1 33867.6 16318.5 0.11 0.11 2.08 9 76.93 42334.2 42335.8 19857.6 0.11 0.11 2.13 10 92.31 50842.7 50843.0 23855.2 0.19 0.19 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3602 to 108268 e- Average count rate measured: 550 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay