*** MIDAS logfile (03SEPpl1.2) *** Fr, 09 Dec 2005 18:37:20 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 09 Dec 2005 18:37:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 051209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.42 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_051209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1683.4 1683.9 822.7 0.11 0.11 2.05 2 6.16 3361.4 3361.8 1622.2 -0.13 -0.13 2.07 3 12.31 6726.3 6725.6 3340.6 -0.12 -0.14 2.01 4 18.47 10091.1 10091.5 4919.3 -0.12 -0.12 2.05 5 27.70 15143.1 15143.1 7451.8 -0.08 -0.08 2.03 6 36.93 20201.7 20201.2 9750.8 -0.03 -0.03 2.07 7 46.16 25258.5 25260.0 12195.2 -0.01 -0.00 2.07 8 61.54 33716.2 33715.8 16221.6 0.11 0.10 2.08 9 76.93 42139.3 42144.1 20000.4 0.09 0.10 2.11 10 92.31 50614.8 50621.4 23641.7 0.18 0.19 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3588 to 107876 e- Average count rate measured: 547 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.24 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay