*** MIDAS logfile (03SEPpl1.2) *** Fr, 04 Nov 2005 15:05:42 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 04 Nov 2005 15:05:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 051104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.72 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_051104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1683.3 1683.7 817.5 0.15 0.14 2.06 2 6.16 3355.2 3355.7 1599.2 -0.22 -0.22 2.10 3 12.31 6720.4 6720.5 3206.0 -0.09 -0.09 2.10 4 18.47 10075.6 10074.8 4919.1 -0.14 -0.15 2.05 5 27.70 15125.0 15124.0 7324.7 -0.06 -0.07 2.06 6 36.93 20174.7 20172.0 9955.3 -0.03 -0.03 2.03 7 46.16 25235.2 25234.0 12091.7 0.04 0.04 2.09 8 61.54 33655.9 33654.2 16177.2 0.06 0.07 2.08 9 76.93 42119.1 42119.1 19784.5 0.18 0.19 2.13 10 92.31 50510.9 50509.5 23430.4 0.12 0.12 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3609 to 108278 e- Average count rate measured: 547 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.28 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay