*** MIDAS logfile (03SEPpl1.2) *** We, 12 Oct 2005 17:27:01 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 12 Oct 2005 17:27:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 051012 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.89 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_051012 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1682.6 1683.0 817.8 0.11 0.12 2.06 2 6.16 3359.9 3360.1 1600.7 -0.14 -0.14 2.10 3 12.31 6722.8 6723.3 3210.0 -0.13 -0.13 2.09 4 18.47 10086.8 10086.4 4912.3 -0.12 -0.12 2.05 5 27.70 15135.5 15137.0 7268.2 -0.09 -0.08 2.08 6 36.93 20191.6 20192.0 9658.9 -0.04 -0.04 2.09 7 46.16 25249.8 25250.5 12189.2 0.00 -0.00 2.07 8 61.54 33700.3 33701.8 16050.9 0.10 0.10 2.10 9 76.93 42128.4 42130.7 19897.0 0.11 0.11 2.12 10 92.31 50592.3 50596.8 23406.7 0.18 0.18 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3609 to 108499 e- Average count rate measured: 547 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.29 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay