*** MIDAS logfile (03SEPpl1.2) *** We, 28 Sep 2005 17:49:32 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 28 Sep 2005 17:49:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050928 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.11 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050928 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1681.9 1682.1 817.6 0.14 0.14 2.06 2 6.16 3355.6 3355.5 1798.4 -0.16 -0.17 1.87 3 12.31 6716.0 6716.5 3262.4 -0.12 -0.12 2.06 4 18.47 10074.2 10073.9 4853.3 -0.13 -0.14 2.08 5 27.70 15117.5 15118.9 7364.3 -0.09 -0.09 2.05 6 36.93 20168.1 20167.6 9789.2 -0.04 -0.04 2.06 7 46.16 25222.2 25223.2 12079.0 0.01 0.01 2.09 8 61.54 33658.5 33660.9 16242.8 0.09 0.10 2.07 9 76.93 42091.3 42092.6 19829.4 0.13 0.13 2.12 10 92.31 50529.1 50532.7 23554.4 0.17 0.17 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3601 to 108174 e- Average count rate measured: 546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.25 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay