*** MIDAS logfile (03SEPpl1.2) *** We, 21 Sep 2005 15:26:29 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 21 Sep 2005 15:26:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050921 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.52 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050921 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1698.6 1698.6 821.2 0.13 0.14 2.07 2 6.16 3387.1 3386.5 1623.1 -0.20 -0.21 2.09 3 12.31 6781.9 6781.4 3218.6 -0.10 -0.10 2.11 4 18.47 10168.7 10167.9 4951.9 -0.14 -0.14 2.05 5 27.70 15263.9 15262.1 7477.1 -0.07 -0.08 2.04 6 36.93 20359.0 20359.0 9861.5 -0.04 -0.03 2.06 7 46.16 25465.8 25465.5 12356.0 0.03 0.03 2.06 8 61.54 33969.5 33968.2 16071.8 0.07 0.08 2.11 9 76.93 42505.5 42504.1 19794.5 0.18 0.18 2.15 10 92.31 50983.3 50982.7 23824.4 0.13 0.13 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3645 to 109397 e- Average count rate measured: 552 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.28 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay