*** MIDAS logfile (03SEPpl1.2) *** Fr, 16 Sep 2005 18:31:52 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 16 Sep 2005 18:32:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050916 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.94 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050916 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1706.7 1706.8 829.7 0.11 0.14 2.06 2 6.16 3407.0 3406.4 1616.5 -0.15 -0.16 2.11 3 12.31 6817.6 6817.8 3241.4 -0.12 -0.12 2.10 4 18.47 10227.7 10228.0 4895.5 -0.12 -0.12 2.09 5 27.70 15347.7 15348.5 7452.7 -0.08 -0.08 2.06 6 36.93 20475.5 20475.5 10065.4 -0.03 -0.04 2.03 7 46.16 25604.0 25604.4 12512.7 0.01 0.00 2.05 8 61.54 34172.3 34174.5 16319.8 0.10 0.10 2.09 9 76.93 42718.0 42722.9 20181.2 0.11 0.11 2.12 10 92.31 51295.5 51298.3 23840.7 0.17 0.17 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3644 to 109537 e- Average count rate measured: 555 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.24 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay