*** MIDAS logfile (03SEPpl1.2) *** We, 07 Sep 2005 13:41:59 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 07 Sep 2005 13:42:16 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050907 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.48 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050907 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1703.7 1703.7 829.7 0.14 0.11 2.05 2 6.16 3399.6 3399.9 1644.1 -0.16 -0.16 2.07 3 12.31 6804.2 6804.9 3356.2 -0.12 -0.11 2.03 4 18.47 10206.0 10206.5 4895.8 -0.13 -0.13 2.08 5 27.70 15316.5 15317.4 7480.3 -0.08 -0.08 2.05 6 36.93 20433.9 20432.5 9950.8 -0.03 -0.03 2.05 7 46.16 25554.2 25554.0 12394.8 0.01 0.02 2.06 8 61.54 34098.8 34099.8 16389.0 0.09 0.10 2.08 9 76.93 42640.6 42639.7 20109.6 0.13 0.13 2.12 10 92.31 51182.2 51181.3 23899.7 0.15 0.16 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3635 to 109202 e- Average count rate measured: 554 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay