*** MIDAS logfile (03SEPpl1.2) *** Th, 25 Aug 2005 18:13:51 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 25 Aug 2005 18:14:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050825 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.42 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050825 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1696.1 1695.9 818.2 0.14 0.13 2.07 2 6.16 3382.6 3383.3 1589.8 -0.19 -0.17 2.13 3 12.31 6772.2 6772.2 3238.5 -0.11 -0.11 2.09 4 18.47 10157.2 10156.6 4914.0 -0.13 -0.13 2.07 5 27.70 15244.2 15243.1 7472.6 -0.07 -0.08 2.04 6 36.93 20334.9 20333.7 9979.3 -0.03 -0.04 2.04 7 46.16 25432.0 25431.4 12425.0 0.02 0.02 2.05 8 61.54 33932.9 33934.4 16315.1 0.09 0.09 2.08 9 76.93 42439.2 42437.5 20064.2 0.14 0.14 2.12 10 92.31 50932.3 50932.2 23876.8 0.15 0.15 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3601 to 108126 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.26 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay