*** MIDAS logfile (03SEPpl1.2) *** Th, 11 Aug 2005 14:29:49 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 11 Aug 2005 14:30:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050811 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.29 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050811 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1688.9 1689.2 828.8 0.13 0.13 2.04 2 6.16 3368.6 3368.4 1611.2 -0.19 -0.19 2.09 3 12.31 6745.7 6744.0 3232.7 -0.09 -0.11 2.09 4 18.47 10114.2 10115.0 4943.7 -0.14 -0.13 2.05 5 27.70 15181.3 15180.5 7408.5 -0.07 -0.08 2.05 6 36.93 20248.9 20249.4 9795.3 -0.04 -0.04 2.07 7 46.16 25327.6 25327.3 12181.3 0.02 0.02 2.08 8 61.54 33790.6 33790.9 16216.7 0.08 0.09 2.08 9 76.93 42270.2 42269.7 19937.6 0.16 0.16 2.12 10 92.31 50716.7 50715.0 23742.4 0.14 0.14 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3599 to 108055 e- Average count rate measured: 549 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay