*** MIDAS logfile (03SEPpl1.2) *** Sa, 06 Aug 2005 19:04:31 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 06 Aug 2005 19:04:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050806 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.86 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050806 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1705.2 1705.2 828.2 0.13 0.13 2.06 2 6.16 3401.3 3400.6 1631.7 -0.19 -0.20 2.08 3 12.31 6810.6 6810.4 3282.3 -0.09 -0.09 2.07 4 18.47 10211.8 10211.2 4923.2 -0.14 -0.14 2.07 5 27.70 15328.3 15327.8 7406.9 -0.07 -0.07 2.07 6 36.93 20446.5 20445.4 9864.2 -0.03 -0.03 2.07 7 46.16 25574.0 25573.3 12269.5 0.03 0.03 2.08 8 61.54 34114.2 34112.0 16222.9 0.08 0.07 2.10 9 76.93 42681.4 42680.1 20296.7 0.16 0.17 2.10 10 92.31 51196.9 51192.8 23825.9 0.12 0.12 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3638 to 109229 e- Average count rate measured: 554 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay