*** MIDAS logfile (03SEPpl1.2) *** We, 27 Jul 2005 14:43:51 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 27 Jul 2005 14:44:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050727 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.05 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050727 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1700.9 1700.7 827.0 0.13 0.12 2.06 2 6.16 3392.2 3392.6 1611.6 -0.18 -0.18 2.10 3 12.31 6791.0 6791.1 3292.9 -0.11 -0.11 2.06 4 18.47 10185.1 10185.1 4882.9 -0.13 -0.13 2.09 5 27.70 15285.9 15285.7 7436.1 -0.08 -0.08 2.06 6 36.93 20391.2 20390.1 9869.8 -0.03 -0.04 2.07 7 46.16 25502.2 25502.0 12504.2 0.02 0.02 2.04 8 61.54 34027.7 34026.9 16461.2 0.09 0.09 2.07 9 76.93 42553.2 42554.7 20302.4 0.13 0.14 2.10 10 92.31 51078.9 51076.1 23551.8 0.16 0.16 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3630 to 109015 e- Average count rate measured: 552 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.03 e-/ADU RON = 4.25 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay