*** MIDAS logfile (03SEPpl1.2) *** Fr, 22 Jul 2005 19:10:03 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 22 Jul 2005 19:10:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050722 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 358.37 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050722 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1715.0 1714.4 826.9 0.14 0.12 2.07 2 6.16 3421.3 3421.6 1636.1 -0.17 -0.16 2.09 3 12.31 6847.7 6847.3 3235.8 -0.12 -0.12 2.12 4 18.47 10271.8 10271.4 4995.9 -0.13 -0.13 2.06 5 27.70 15415.7 15415.5 7523.5 -0.08 -0.08 2.05 6 36.93 20565.2 20564.3 10065.8 -0.03 -0.03 2.04 7 46.16 25718.2 25718.0 12448.0 0.01 0.01 2.07 8 61.54 34320.8 34320.8 16386.3 0.10 0.10 2.09 9 76.93 42913.1 42915.6 20436.9 0.12 0.13 2.10 10 92.31 51517.7 51519.2 23792.4 0.17 0.17 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3669 to 110231 e- Average count rate measured: 557 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.26 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay