*** MIDAS logfile (03SEPpl1.2) *** We, 06 Jul 2005 17:18:43 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 06 Jul 2005 17:19:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050706 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.26 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050706 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1712.1 1712.7 827.0 0.17 0.17 2.07 2 6.16 3411.8 3412.4 1641.1 -0.23 -0.23 2.08 3 12.31 6835.4 6834.7 3286.4 -0.08 -0.09 2.08 4 18.47 10248.0 10247.0 4986.9 -0.14 -0.15 2.05 5 27.70 15383.3 15382.2 7411.7 -0.07 -0.07 2.08 6 36.93 20518.1 20516.9 9880.8 -0.04 -0.04 2.08 7 46.16 25669.9 25666.5 12361.3 0.04 0.04 2.08 8 61.54 34232.3 34229.8 16171.6 0.06 0.06 2.12 9 76.93 42846.6 42842.4 20168.6 0.19 0.19 2.12 10 92.31 51371.4 51369.0 23738.5 0.10 0.11 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3686 to 110589 e- Average count rate measured: 556 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.30 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay