*** MIDAS logfile (03SEPpl1.2) *** We, 29 Jun 2005 21:04:00 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 29 Jun 2005 21:04:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050629 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.20 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050629 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1722.6 1722.8 828.4 0.12 0.13 2.08 2 6.16 3437.6 3437.9 1618.0 -0.16 -0.16 2.12 3 12.31 6880.1 6880.0 3310.9 -0.12 -0.12 2.08 4 18.47 10321.1 10321.0 4968.4 -0.12 -0.13 2.08 5 27.70 15489.9 15488.5 7558.3 -0.07 -0.08 2.05 6 36.93 20661.3 20662.2 10016.4 -0.04 -0.04 2.06 7 46.16 25841.0 25840.8 12418.4 0.02 0.01 2.08 8 61.54 34481.5 34483.7 16580.7 0.09 0.10 2.08 9 76.93 43119.7 43121.4 20410.9 0.13 0.13 2.11 10 92.31 51758.5 51760.5 23982.4 0.16 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3683 to 110647 e- Average count rate measured: 560 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.25 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay