*** MIDAS logfile (03SEPpl1.2) *** Sa, 18 Jun 2005 13:20:17 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 18 Jun 2005 13:20:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050618 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.07 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050618 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1732.2 1731.9 830.3 0.14 0.13 2.09 2 6.16 3456.2 3457.1 1649.2 -0.16 -0.15 2.10 3 12.31 6918.6 6918.5 3351.5 -0.12 -0.12 2.06 4 18.47 10378.3 10378.2 5056.7 -0.13 -0.13 2.05 5 27.70 15575.0 15575.8 7592.9 -0.08 -0.08 2.05 6 36.93 20777.8 20778.3 10103.6 -0.03 -0.04 2.06 7 46.16 25984.9 25985.5 12495.8 0.01 0.01 2.08 8 61.54 34675.5 34676.3 16514.7 0.09 0.09 2.10 9 76.93 43355.1 43359.3 20292.7 0.12 0.12 2.14 10 92.31 52050.1 52051.5 24104.9 0.16 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3729 to 112053 e- Average count rate measured: 563 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.33 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay