*** MIDAS logfile (03SEPpl1.2) *** Tu, 14 Jun 2005 08:20:54 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Tu, 14 Jun 2005 08:21:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050614 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.87 stddev = 2.05 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050614 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1715.9 1716.2 860.1 0.12 0.13 1.99 2 6.16 3423.6 3424.3 1608.8 -0.17 -0.16 2.13 3 12.31 6853.5 6853.3 3317.1 -0.11 -0.11 2.07 4 18.47 10279.5 10279.0 4979.6 -0.12 -0.13 2.06 5 27.70 15427.4 15427.4 7497.7 -0.07 -0.08 2.06 6 36.93 20578.3 20578.9 9997.0 -0.04 -0.04 2.06 7 46.16 25736.5 25737.2 12448.1 0.01 0.01 2.07 8 61.54 34341.6 34340.9 16337.2 0.09 0.08 2.10 9 76.93 42948.2 42950.8 20010.4 0.14 0.14 2.15 10 92.31 51544.6 51544.6 24235.8 0.15 0.15 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3666 to 110123 e- Average count rate measured: 558 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.38 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay