*** MIDAS logfile (03SEPpl1.2) *** Fr, 03 Jun 2005 14:30:54 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 03 Jun 2005 14:31:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050603 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.74 stddev = 2.03 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050603 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1740.1 1740.0 833.9 0.13 0.13 2.09 2 6.16 3470.6 3470.5 1656.3 -0.18 -0.18 2.10 3 12.31 6947.4 6946.7 3375.9 -0.11 -0.11 2.06 4 18.47 10420.3 10419.1 5074.4 -0.13 -0.13 2.05 5 27.70 15637.3 15637.0 7575.5 -0.08 -0.08 2.06 6 36.93 20861.5 20860.9 10060.2 -0.03 -0.03 2.07 7 46.16 26091.2 26088.5 12573.9 0.02 0.02 2.08 8 61.54 34812.7 34811.3 16796.9 0.09 0.09 2.07 9 76.93 43534.6 43534.3 20409.2 0.13 0.14 2.13 10 92.31 52250.7 52251.5 24200.0 0.15 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3734 to 112126 e- Average count rate measured: 565 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.36 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay