*** MIDAS logfile (03SEPpl1.2) *** Su, 29 May 2005 17:31:52 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 29 May 2005 17:32:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050529 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.97 stddev = 2.03 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050529 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1744.1 1744.2 827.1 0.13 0.14 2.11 2 6.16 3478.8 3478.8 1655.8 -0.18 -0.18 2.10 3 12.31 6965.4 6964.8 3333.2 -0.10 -0.10 2.09 4 18.47 10444.3 10443.6 5033.9 -0.14 -0.14 2.07 5 27.70 15677.0 15676.1 7614.4 -0.07 -0.07 2.06 6 36.93 20911.5 20910.1 10071.8 -0.03 -0.04 2.08 7 46.16 26157.0 26154.5 12615.7 0.03 0.03 2.07 8 61.54 34892.7 34890.2 16868.4 0.08 0.08 2.07 9 76.93 43652.8 43648.9 20438.3 0.16 0.16 2.14 10 92.31 52363.4 52362.5 24398.1 0.12 0.13 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3725 to 111846 e- Average count rate measured: 567 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.33 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay