*** MIDAS logfile (03SEPpl1.2) *** Sa, 30 Apr 2005 15:49:28 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 30 Apr 2005 15:49:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050430 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.18 stddev = 2.03 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050430 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1750.5 1750.0 843.8 0.15 0.16 2.07 2 6.16 3489.2 3489.1 1687.2 -0.23 -0.22 2.07 3 12.31 6990.6 6990.1 3322.0 -0.08 -0.08 2.10 4 18.47 10479.2 10479.0 5095.2 -0.15 -0.14 2.06 5 27.70 15731.1 15730.9 7695.1 -0.07 -0.07 2.04 6 36.93 20982.6 20981.0 10036.0 -0.04 -0.04 2.09 7 46.16 26249.1 26247.4 12694.2 0.04 0.04 2.07 8 61.54 35009.2 35007.0 16794.6 0.07 0.07 2.08 9 76.93 43816.1 43812.9 20905.8 0.19 0.19 2.10 10 92.31 52532.8 52531.2 24428.3 0.10 0.10 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3726 to 111826 e- Average count rate measured: 569 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.32 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay