*** MIDAS logfile (03SEPpl1.2) *** Sa, 16 Apr 2005 17:25:11 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 16 Apr 2005 17:25:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050416 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.45 stddev = 2.03 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050416 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1734.2 1733.6 828.5 0.13 0.14 2.09 2 6.16 3459.8 3459.0 1642.0 -0.17 -0.17 2.11 3 12.31 6926.2 6925.1 3357.7 -0.10 -0.11 2.06 4 18.47 10386.4 10386.4 5055.8 -0.14 -0.14 2.05 5 27.70 15591.5 15589.5 7553.8 -0.07 -0.08 2.06 6 36.93 20796.8 20796.2 10086.5 -0.03 -0.03 2.06 7 46.16 26010.9 26010.5 12551.8 0.02 0.02 2.07 8 61.54 34702.0 34701.3 16668.6 0.08 0.08 2.08 9 76.93 43404.6 43405.6 20551.1 0.14 0.15 2.11 10 92.31 52080.6 52081.3 24178.4 0.13 0.13 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3704 to 111268 e- Average count rate measured: 563 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.34 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay