*** MIDAS logfile (03SEPpl1.2) *** Su, 10 Apr 2005 22:33:58 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 10 Apr 2005 22:34:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050410 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.65 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050410 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1758.1 1758.4 851.0 0.10 0.11 2.07 2 6.16 3512.4 3512.7 1676.9 -0.10 -0.10 2.09 3 12.31 7024.4 7025.5 3864.2 -0.14 -0.14 1.82 4 18.47 10542.5 10541.9 5055.1 -0.11 -0.12 2.09 5 27.70 15818.0 15819.5 7753.5 -0.08 -0.09 2.04 6 36.93 21103.6 21104.4 10259.2 -0.03 -0.03 2.06 7 46.16 26386.4 26389.1 12813.0 -0.01 -0.01 2.06 8 61.54 35226.3 35227.8 17039.2 0.12 0.11 2.07 9 76.93 44011.6 44020.2 20873.5 0.07 0.08 2.11 10 92.31 52878.7 52888.2 24811.5 0.19 0.20 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3745 to 112646 e- Average count rate measured: 572 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.27 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay