*** MIDAS logfile (03SEPpl1.2) *** Sa, 09 Apr 2005 18:15:08 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 09 Apr 2005 18:15:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050409 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.38 stddev = 2.03 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050409 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1744.6 1745.0 831.6 0.13 0.11 2.10 2 6.16 3482.2 3483.2 1688.5 -0.16 -0.15 2.06 3 12.31 6969.2 6970.0 3372.8 -0.12 -0.12 2.07 4 18.47 10455.4 10455.8 4994.7 -0.12 -0.12 2.09 5 27.70 15690.6 15690.8 7688.0 -0.08 -0.08 2.04 6 36.93 20930.6 20932.1 10154.0 -0.04 -0.03 2.06 7 46.16 26176.5 26177.6 12761.5 0.01 0.01 2.05 8 61.54 34932.0 34934.1 16647.1 0.10 0.10 2.10 9 76.93 43674.1 43676.7 20670.9 0.12 0.12 2.11 10 92.31 52434.2 52434.8 24423.1 0.16 0.16 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3724 to 111922 e- Average count rate measured: 567 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.33 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay