*** MIDAS logfile (03SEPpl1.2) *** Su, 13 Mar 2005 15:41:12 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 13 Mar 2005 15:41:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050313 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.46 stddev = 2.02 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050313 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1766.2 1766.0 854.1 0.12 0.13 2.07 2 6.16 3523.3 3522.8 1662.2 -0.17 -0.17 2.12 3 12.31 7052.3 7052.7 3366.1 -0.11 -0.10 2.10 4 18.47 10578.4 10578.4 5162.6 -0.12 -0.12 2.05 5 27.70 15875.8 15875.6 7753.6 -0.07 -0.07 2.05 6 36.93 21177.2 21178.2 10204.2 -0.03 -0.03 2.08 7 46.16 26486.6 26486.6 12841.2 0.03 0.02 2.06 8 61.54 35342.5 35342.2 16696.9 0.10 0.10 2.12 9 76.93 44198.5 44196.2 20884.6 0.15 0.14 2.12 10 92.31 53007.0 53009.5 24537.5 0.09 0.09 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3791 to 113799 e- Average count rate measured: 574 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.33 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay