*** MIDAS logfile (03SEPpl1.2) *** Fr, 04 Mar 2005 20:47:58 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 04 Mar 2005 20:48:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050304 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.23 stddev = 2.05 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050304 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1804.9 1805.2 878.5 0.13 0.11 2.05 2 6.16 3603.2 3603.5 1701.0 -0.14 -0.14 2.12 3 12.31 7210.6 7210.7 3465.2 -0.12 -0.13 2.08 4 18.47 10817.1 10817.8 5228.7 -0.12 -0.12 2.07 5 27.70 16232.4 16232.8 7884.9 -0.08 -0.08 2.06 6 36.93 21656.2 21655.7 10417.0 -0.03 -0.03 2.08 7 46.16 27081.1 27082.8 13116.4 0.01 0.01 2.06 8 61.54 36141.7 36144.2 17315.9 0.10 0.10 2.09 9 76.93 45174.6 45179.7 21210.3 0.09 0.10 2.13 10 92.31 54250.3 54257.5 25064.1 0.17 0.18 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3878 to 116570 e- Average count rate measured: 587 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.40 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay