*** MIDAS logfile (03SEPpl1.2) *** Mo, 28 Feb 2005 15:44:30 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 28 Feb 2005 15:44:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050228 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.89 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050228 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1630.0 1629.4 840.6 0.12 0.10 1.94 2 6.16 3249.3 3249.3 1661.6 -0.16 -0.16 1.96 3 12.31 6502.7 6501.8 3251.9 -0.08 -0.09 2.00 4 18.47 9750.9 9751.7 4862.9 -0.10 -0.09 2.01 5 27.70 14633.4 14632.4 7250.5 -0.05 -0.05 2.02 6 36.93 19517.4 19518.0 9667.6 -0.01 -0.01 2.02 7 46.16 24407.7 24407.2 12056.7 0.03 0.03 2.02 8 61.54 32560.6 32561.0 15658.3 0.08 0.09 2.08 9 76.93 40706.7 40707.2 18936.4 0.10 0.10 2.15 10 92.31 48835.4 48836.7 22707.8 0.07 0.08 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3517 to 105404 e- Average count rate measured: 529 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.03 e-/ADU RON = 4.38 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay