*** MIDAS logfile (03SEPpl1.2) *** Sa, 12 Feb 2005 16:43:09 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 12 Feb 2005 16:43:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050212 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.70 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1627.6 1627.5 761.9 0.10 0.12 2.14 2 6.16 3250.7 3251.4 1523.9 -0.13 -0.11 2.13 3 12.31 6504.6 6504.2 3077.5 -0.13 -0.14 2.11 4 18.47 9759.0 9758.5 4738.3 -0.12 -0.13 2.06 5 27.70 14644.4 14644.3 6998.6 -0.09 -0.09 2.09 6 36.93 19537.2 19538.4 9437.8 -0.04 -0.04 2.07 7 46.16 24431.1 24431.1 11845.5 -0.00 -0.01 2.06 8 61.54 32609.4 32612.8 15656.1 0.10 0.11 2.08 9 76.93 40761.7 40765.1 19200.1 0.10 0.10 2.12 10 92.31 48960.6 48964.8 22770.2 0.20 0.19 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3472 to 104462 e- Average count rate measured: 529 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.25 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay