*** MIDAS logfile (03SEPpl1.2) *** Sa, 22 Jan 2005 22:11:45 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 22 Jan 2005 22:12:01 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050122 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.24 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050122 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1663.7 1663.8 784.4 0.13 0.13 2.12 2 6.16 3321.3 3320.5 1566.0 -0.15 -0.16 2.12 3 12.31 6647.1 6647.3 3131.6 -0.12 -0.12 2.12 4 18.47 9972.9 9972.6 4784.1 -0.11 -0.12 2.08 5 27.70 14965.2 14965.8 7239.6 -0.08 -0.08 2.07 6 36.93 19965.4 19966.0 9650.7 -0.03 -0.02 2.07 7 46.16 24966.5 24966.5 11946.1 0.01 0.01 2.09 8 61.54 33319.5 33320.0 15975.2 0.10 0.10 2.09 9 76.93 41652.4 41654.4 19662.2 0.10 0.11 2.12 10 92.31 50008.9 50008.7 23268.8 0.15 0.15 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3549 to 106661 e- Average count rate measured: 541 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.27 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay