*** MIDAS logfile (03SEPpl1.2) *** Su, 02 Jan 2005 16:01:17 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 02 Jan 2005 16:01:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050102 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.04 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050102 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1659.3 1659.2 782.3 0.11 0.10 2.12 2 6.16 3314.7 3315.1 1547.1 -0.11 -0.10 2.14 3 12.31 6629.9 6630.2 3137.2 -0.15 -0.16 2.11 4 18.47 9951.9 9952.5 4764.2 -0.10 -0.11 2.09 5 27.70 14929.5 14932.0 7203.2 -0.10 -0.09 2.07 6 36.93 19921.9 19923.3 9576.4 -0.03 -0.03 2.08 7 46.16 24904.5 24907.5 12005.3 -0.02 -0.02 2.07 8 61.54 33256.6 33259.5 15766.5 0.13 0.13 2.11 9 76.93 41541.2 41548.1 19419.7 0.05 0.06 2.14 10 92.31 49932.0 49939.5 23361.8 0.22 0.22 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3545 to 106674 e- Average count rate measured: 540 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.23 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay