*** MIDAS logfile (03SEPpl1.2) *** Su, 26 Dec 2004 17:22:29 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 26 Dec 2004 17:22:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041226 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.38 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041226 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1657.3 1656.8 788.8 0.10 0.11 2.10 2 6.16 3310.8 3310.4 1556.3 -0.09 -0.10 2.13 3 12.31 6620.3 6619.7 3125.6 -0.16 -0.17 2.12 4 18.47 9935.0 9936.3 4776.3 -0.12 -0.12 2.08 5 27.70 14906.8 14907.5 7206.3 -0.10 -0.11 2.07 6 36.93 19888.6 19891.4 9601.7 -0.04 -0.04 2.07 7 46.16 24866.5 24869.2 11922.4 -0.02 -0.03 2.09 8 61.54 33205.3 33210.7 15870.3 0.12 0.13 2.09 9 76.93 41489.6 41493.0 19426.9 0.08 0.08 2.14 10 92.31 49867.5 49882.3 23239.8 0.24 0.26 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3544 to 106664 e- Average count rate measured: 539 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.23 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay