*** MIDAS logfile (03SEPpl1.2) *** Su, 19 Dec 2004 13:16:09 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 19 Dec 2004 13:16:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041219 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.00 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041219 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1662.7 1662.9 786.0 0.14 0.13 2.12 2 6.16 3316.3 3316.8 1568.6 -0.20 -0.19 2.11 3 12.31 6641.0 6641.0 3168.1 -0.10 -0.10 2.10 4 18.47 9958.9 9959.1 4757.7 -0.14 -0.14 2.09 5 27.70 14947.8 14946.7 7252.1 -0.08 -0.08 2.06 6 36.93 19938.7 19938.4 9535.0 -0.04 -0.04 2.09 7 46.16 24939.4 24939.9 12101.8 0.02 0.02 2.06 8 61.54 33273.2 33273.6 15991.3 0.08 0.08 2.08 9 76.93 41627.1 41625.5 19482.8 0.17 0.16 2.14 10 92.31 49942.3 49944.2 23648.4 0.15 0.15 2.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3519 to 105682 e- Average count rate measured: 540 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay