*** MIDAS logfile (03SEPpl1.2) *** We, 08 Dec 2004 19:58:34 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 08 Dec 2004 19:58:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041208 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.45 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041208 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1663.1 1663.6 786.2 0.11 0.12 2.12 2 6.16 3323.0 3322.2 1561.9 -0.10 -0.14 2.13 3 12.31 6647.2 6646.8 3119.8 -0.14 -0.17 2.13 4 18.47 9974.3 9974.4 4822.4 -0.12 -0.14 2.07 5 27.70 14964.7 14966.3 7206.6 -0.11 -0.12 2.08 6 36.93 19967.7 19968.3 9676.9 -0.04 -0.06 2.06 7 46.16 24964.7 24967.6 12103.0 -0.02 -0.04 2.06 8 61.54 33331.0 33335.3 15854.0 0.11 0.09 2.10 9 76.93 41652.3 41653.7 19442.5 0.08 0.05 2.14 10 92.31 50059.7 50167.8 23295.9 0.23 0.42 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3567 to 107460 e- Average count rate measured: 541 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.29 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.011 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.32/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay