*** MIDAS logfile (03SEPpl1.2) *** Tu, 07 Dec 2004 15:25:54 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Tu, 07 Dec 2004 15:26:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041207 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.70 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041207 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1651.3 1651.8 777.0 0.14 0.15 2.13 2 6.16 3294.0 3293.7 1559.7 -0.23 -0.24 2.11 3 12.31 6598.6 6598.4 3091.1 -0.11 -0.12 2.13 4 18.47 9901.1 9901.2 4771.9 -0.09 -0.10 2.07 5 27.70 14856.9 14859.3 7128.1 -0.06 -0.05 2.08 6 36.93 19820.5 19822.8 9644.9 -0.01 -0.00 2.06 7 46.16 24785.5 24786.1 12034.9 0.03 0.02 2.06 8 61.54 33084.8 33085.3 15926.3 0.14 0.13 2.08 9 76.93 41348.1 41352.2 19640.7 0.12 0.12 2.11 10 92.31 49599.2 49604.1 22904.8 0.08 0.08 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3525 to 105862 e- Average count rate measured: 537 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.23 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay