*** MIDAS logfile (03SEPpl1.2) *** Mo, 06 Dec 2004 18:22:23 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 06 Dec 2004 18:22:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041206 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.72 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041206 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1650.1 1650.6 772.2 0.12 0.12 2.14 2 6.16 3296.4 3296.1 1548.9 -0.12 -0.14 2.13 3 12.31 6594.1 6594.8 3101.7 -0.14 -0.14 2.13 4 18.47 9895.9 9895.9 4770.5 -0.12 -0.12 2.07 5 27.70 14848.5 14848.8 7244.2 -0.09 -0.09 2.05 6 36.93 19810.9 19813.2 9522.9 -0.04 -0.03 2.08 7 46.16 24771.8 24773.4 11974.3 -0.01 -0.00 2.07 8 61.54 33069.1 33069.7 15894.8 0.11 0.11 2.08 9 76.93 41326.6 41332.0 19369.9 0.09 0.10 2.13 10 92.31 49648.5 49652.2 22980.3 0.20 0.20 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3538 to 106427 e- Average count rate measured: 537 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.26 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay