*** MIDAS logfile (03SEPpl1.2) *** Su, 28 Nov 2004 17:16:19 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 28 Nov 2004 17:16:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041128 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.34 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041128 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1662.6 1662.0 773.6 0.12 0.14 2.15 2 6.16 3317.9 3316.9 1562.7 -0.16 -0.16 2.12 3 12.31 6639.7 6639.3 3170.9 -0.13 -0.13 2.09 4 18.47 9960.4 9960.2 4740.9 -0.13 -0.13 2.10 5 27.70 14948.0 14948.1 7235.3 -0.08 -0.08 2.07 6 36.93 19942.1 19941.8 9540.8 -0.03 -0.04 2.09 7 46.16 24937.6 24938.3 11939.8 0.01 0.00 2.09 8 61.54 33281.2 33283.7 15904.4 0.10 0.10 2.09 9 76.93 41613.8 41615.5 19661.8 0.13 0.12 2.12 10 92.31 49960.3 49963.3 23344.9 0.17 0.17 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3537 to 106317 e- Average count rate measured: 540 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.19 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay