*** MIDAS logfile (03SEPpl1.2) *** Sa, 20 Nov 2004 19:27:27 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 20 Nov 2004 19:27:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041120 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.82 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041120 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1670.1 1669.6 781.1 0.14 0.12 2.14 2 6.16 3330.9 3331.0 1571.8 -0.19 -0.18 2.12 3 12.31 6669.4 6668.6 3144.4 -0.11 -0.12 2.12 4 18.47 10002.2 10002.7 4713.3 -0.14 -0.13 2.12 5 27.69 15012.6 15012.9 7221.8 -0.08 -0.08 2.08 6 36.93 20026.0 20025.8 9598.2 -0.04 -0.04 2.09 7 46.16 25048.3 25047.2 12098.1 0.02 0.02 2.07 8 61.54 33420.7 33421.7 15841.6 0.09 0.09 2.11 9 76.93 41798.4 41798.7 19778.5 0.14 0.14 2.11 10 92.31 50167.3 50169.1 23224.4 0.16 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3571 to 107280 e- Average count rate measured: 543 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.20 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay