*** MIDAS logfile (03SEPpl1.2) *** Sa, 13 Nov 2004 23:12:29 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 13 Nov 2004 23:12:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041113 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.10 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041113 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1697.1 1696.4 790.0 0.12 0.12 2.15 2 6.16 3388.0 3387.7 1610.2 -0.14 -0.14 2.10 3 12.31 6780.7 6780.5 3231.5 -0.12 -0.12 2.10 4 18.47 10173.5 10172.4 4849.1 -0.12 -0.13 2.10 5 27.70 15267.5 15267.2 7409.8 -0.08 -0.08 2.06 6 36.93 20367.3 20365.6 9756.8 -0.03 -0.04 2.09 7 46.16 25469.7 25471.1 12135.7 0.01 0.01 2.10 8 61.54 33989.6 33992.1 16135.3 0.09 0.10 2.11 9 76.93 42489.4 42496.2 19946.5 0.10 0.11 2.13 10 92.31 51023.1 51027.6 23789.9 0.17 0.17 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3630 to 109151 e- Average count rate measured: 552 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.19 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay