*** MIDAS logfile (03SEPpl1.2) *** Fr, 01 Oct 2004 15:58:02 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 01 Oct 2004 15:58:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041001 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.81 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_041001 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1576.3 1576.7 726.3 0.13 0.12 2.17 2 6.16 3148.0 3148.3 1479.6 -0.13 -0.14 2.13 3 12.31 6299.8 6300.3 2979.7 -0.13 -0.13 2.11 4 18.47 9452.4 9452.0 4588.8 -0.12 -0.13 2.06 5 27.70 14183.5 14183.4 6809.0 -0.10 -0.10 2.08 6 36.93 18922.3 18923.6 9173.6 -0.04 -0.04 2.06 7 46.16 23662.5 23663.1 11521.6 -0.01 -0.00 2.05 8 61.54 31587.0 31588.4 15181.8 0.11 0.11 2.08 9 76.93 39481.3 39483.9 18597.6 0.10 0.11 2.12 10 92.31 47422.9 47429.8 22065.5 0.19 0.21 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3361 to 101119 e- Average count rate measured: 513 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.19 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay