*** MIDAS logfile (03SEPpl1.2) *** Fr, 24 Sep 2004 20:09:23 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 24 Sep 2004 20:09:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040924 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.60 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040924 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1637.3 1637.1 764.5 0.11 0.13 2.14 2 6.16 3269.9 3269.8 1536.9 -0.13 -0.13 2.13 3 12.31 6542.8 6543.2 3083.8 -0.13 -0.12 2.12 4 18.47 9816.6 9816.1 4703.1 -0.12 -0.13 2.09 5 27.70 14729.2 14730.1 7132.1 -0.09 -0.09 2.07 6 36.93 19653.2 19653.5 9416.8 -0.03 -0.03 2.09 7 46.16 24573.6 24573.7 11841.2 -0.00 -0.01 2.08 8 61.54 32800.1 32802.2 15812.0 0.10 0.10 2.07 9 76.93 40997.2 41003.5 19313.9 0.09 0.10 2.12 10 92.31 49240.6 49246.3 22956.7 0.18 0.18 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3485 to 104830 e- Average count rate measured: 533 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay