*** MIDAS logfile (03SEPpl1.2) *** Sa, 07 Aug 2004 21:23:09 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 07 Aug 2004 21:23:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040807 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.49 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040807 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1653.7 1654.1 776.8 0.14 0.15 2.13 2 6.16 3300.6 3300.0 1541.2 -0.16 -0.18 2.14 3 12.31 6608.8 6607.7 3102.2 -0.10 -0.11 2.13 4 18.47 9910.9 9911.3 4799.1 -0.13 -0.13 2.07 5 27.70 14874.5 14875.2 7146.7 -0.08 -0.08 2.08 6 36.93 19842.6 19843.4 9525.5 -0.04 -0.04 2.08 7 46.16 24817.2 24817.5 11926.1 0.01 0.01 2.08 8 61.54 33114.9 33115.8 15674.2 0.09 0.09 2.11 9 76.93 41414.3 41415.8 19538.2 0.13 0.14 2.12 10 92.31 49703.7 49704.8 23340.8 0.15 0.15 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3515 to 105647 e- Average count rate measured: 538 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.19 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay