*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Jul 2004 13:28:30 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Jul 2004 13:28:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040720 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.24 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1679.0 1679.2 797.7 0.15 0.15 2.10 2 6.16 3349.4 3349.5 1584.4 -0.19 -0.18 2.11 3 12.31 6707.0 6707.1 3186.9 -0.10 -0.10 2.10 4 18.47 10058.4 10058.9 4832.7 -0.14 -0.13 2.08 5 27.70 15096.1 15096.7 7372.7 -0.09 -0.08 2.05 6 36.93 20140.6 20139.2 9906.0 -0.03 -0.04 2.03 7 46.16 25189.2 25188.5 12139.5 0.02 0.02 2.07 8 61.54 33609.9 33606.8 15998.9 0.09 0.08 2.10 9 76.93 42033.5 42033.4 19837.3 0.14 0.14 2.12 10 92.31 50442.3 50444.7 23833.0 0.14 0.15 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3556 to 106836 e- Average count rate measured: 546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.18 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay