*** MIDAS logfile (vers. 02SEPpl1.3) *** Sa, 03 Jul 2004 17:33:01 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Sa, 03 Jul 2004 17:33:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040703 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.99 stddev = 1.95 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040703 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1680.4 1679.8 793.7 0.12 0.12 2.12 2 6.16 3353.5 3353.3 1582.1 -0.17 -0.16 2.12 3 12.31 6712.2 6712.9 3210.2 -0.13 -0.12 2.09 4 18.47 10069.5 10069.7 4771.6 -0.13 -0.13 2.11 5 27.70 15111.6 15112.4 7199.6 -0.08 -0.08 2.10 6 36.93 20160.7 20160.8 9564.8 -0.03 -0.03 2.11 7 46.16 25214.9 25213.3 12159.0 0.02 0.01 2.07 8 61.54 33647.2 33646.0 16136.3 0.10 0.09 2.09 9 76.93 42072.1 42072.7 19957.7 0.13 0.13 2.11 10 92.31 50503.6 50505.8 23320.1 0.16 0.16 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3590 to 107910 e- Average count rate measured: 546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.18 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay