*** MIDAS logfile (vers. 02SEPpl1.3) *** Th, 17 Jun 2004 13:38:48 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Th, 17 Jun 2004 13:39:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040617 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.79 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040617 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1614.6 1615.0 751.3 0.11 0.12 2.15 2 6.16 3224.9 3224.6 1501.0 -0.14 -0.14 2.15 3 12.31 6453.6 6453.5 3078.2 -0.12 -0.13 2.10 4 18.47 9682.5 9682.9 4574.6 -0.12 -0.12 2.12 5 27.70 14530.1 14529.9 6989.7 -0.08 -0.09 2.08 6 36.93 19384.0 19385.5 9383.7 -0.04 -0.03 2.07 7 46.16 24241.3 24241.2 11795.0 0.01 0.00 2.06 8 61.54 32354.4 32355.4 15530.7 0.10 0.10 2.08 9 76.93 40443.5 40449.1 19195.3 0.10 0.11 2.11 10 92.31 48569.2 48573.0 22752.3 0.18 0.18 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3418 to 102807 e- Average count rate measured: 525 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.16 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay