*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 16 May 2004 14:41:29 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 16 May 2004 14:41:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040516 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.84 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040516 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1698.2 1698.0 785.0 0.18 0.17 2.16 2 6.16 3382.6 3382.2 1585.2 -0.27 -0.27 2.13 3 12.31 6779.3 6779.6 3196.8 -0.08 -0.06 2.12 4 18.47 10162.4 10160.7 4819.4 -0.15 -0.16 2.11 5 27.70 15257.8 15255.9 7374.2 -0.06 -0.06 2.07 6 36.93 20349.4 20345.8 9857.8 -0.04 -0.04 2.06 7 46.16 25459.8 25457.1 12240.8 0.05 0.06 2.08 8 61.54 33948.6 33943.2 16194.0 0.06 0.06 2.10 9 76.93 42508.4 42500.0 19959.2 0.23 0.22 2.13 10 92.31 50936.7 50930.3 23803.7 0.08 0.08 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3618 to 108529 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.27 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay