*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 30 Apr 2004 15:53:29 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 30 Apr 2004 15:53:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040430 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.81 stddev = 1.95 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040430 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1725.8 1725.6 807.0 0.19 0.20 2.14 2 6.16 3433.1 3432.4 1603.6 -0.35 -0.35 2.14 3 12.31 6888.3 6887.3 3270.5 -0.03 -0.02 2.11 4 18.47 10316.7 10314.9 4974.4 -0.18 -0.18 2.07 5 27.70 15494.5 15491.1 7585.5 -0.05 -0.05 2.04 6 36.93 20662.7 20659.3 9927.5 -0.04 -0.03 2.08 7 46.16 25863.0 25857.6 12612.6 0.10 0.10 2.05 8 61.54 34458.0 34452.9 16617.3 0.03 0.03 2.07 9 76.93 43198.9 43183.9 20257.6 0.32 0.30 2.13 10 92.31 51682.9 51672.4 24045.3 0.02 0.01 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3684 to 110328 e- Average count rate measured: 560 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.17 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.31/-0.35 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay