*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Apr 2004 17:11:48 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Apr 2004 17:12:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040420 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.14 stddev = 1.95 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040420 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1727.1 1727.5 806.8 0.12 0.13 2.14 2 6.16 3447.2 3447.4 1626.7 -0.18 -0.18 2.12 3 12.31 6902.2 6902.5 3269.0 -0.11 -0.11 2.11 4 18.47 10352.7 10352.9 5004.8 -0.14 -0.14 2.07 5 27.70 15539.1 15538.7 7662.4 -0.08 -0.08 2.03 6 36.93 20730.1 20729.2 10018.6 -0.03 -0.03 2.07 7 46.16 25927.3 25928.7 12330.3 0.02 0.03 2.10 8 61.54 34595.4 34593.7 16402.4 0.10 0.09 2.11 9 76.93 43265.3 43265.0 20430.0 0.14 0.14 2.12 10 92.31 51922.1 51922.4 24255.6 0.15 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3687 to 110837 e- Average count rate measured: 562 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.17 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay