*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 12 Apr 2004 16:43:35 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 12 Apr 2004 16:43:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040412 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.68 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040412 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1726.0 1725.9 821.4 0.12 0.11 2.10 2 6.16 3447.4 3447.2 1621.2 -0.12 -0.12 2.13 3 12.31 6897.3 6897.4 3229.6 -0.14 -0.14 2.14 4 18.47 10350.1 10349.7 4965.3 -0.12 -0.12 2.08 5 27.70 15529.6 15530.0 7478.5 -0.10 -0.09 2.08 6 36.93 20720.5 20721.0 10097.5 -0.03 -0.03 2.05 7 46.16 25908.7 25909.9 12468.5 -0.01 -0.00 2.08 8 61.54 34585.2 34589.4 16457.5 0.11 0.12 2.10 9 76.93 43223.7 43228.7 20142.5 0.09 0.10 2.15 10 92.31 51922.2 51923.3 23873.5 0.19 0.19 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3727 to 112119 e- Average count rate measured: 561 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.24 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay