*** MIDAS logfile (vers. 02SEPpl1.3) *** We, 07 Apr 2004 19:03:40 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** We, 07 Apr 2004 19:03:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040407 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.45 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040407 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1723.8 1723.9 797.5 0.12 0.11 2.16 2 6.16 3441.1 3441.9 1625.7 -0.15 -0.14 2.12 3 12.31 6887.1 6887.7 3277.6 -0.13 -0.13 2.10 4 18.47 10333.0 10333.5 5022.9 -0.12 -0.12 2.06 5 27.70 15506.9 15507.3 7545.4 -0.08 -0.09 2.06 6 36.93 20687.6 20687.3 9944.8 -0.03 -0.04 2.08 7 46.16 25870.5 25870.6 12435.2 0.01 0.00 2.08 8 61.54 34530.0 34532.5 16317.1 0.11 0.11 2.12 9 76.93 43158.5 43165.8 19996.9 0.10 0.11 2.16 10 92.31 51832.4 51838.5 24227.0 0.18 0.18 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3703 to 111357 e- Average count rate measured: 561 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.24 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay