*** MIDAS logfile (vers. 02SEPpl1.3) *** We, 24 Mar 2004 17:48:55 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** We, 24 Mar 2004 17:49:13 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040324 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.26 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040324 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1740.1 1740.2 808.3 0.12 0.12 2.15 2 6.16 3473.2 3472.8 1627.7 -0.16 -0.17 2.13 3 12.31 6951.3 6951.5 3294.0 -0.12 -0.12 2.11 4 18.47 10428.3 10427.2 4977.2 -0.13 -0.13 2.10 5 27.70 15649.3 15650.2 7567.3 -0.09 -0.08 2.07 6 36.93 20879.2 20877.3 10112.6 -0.03 -0.03 2.06 7 46.16 26112.0 26112.3 12526.5 0.02 0.02 2.08 8 61.54 34845.0 34845.5 16395.1 0.10 0.10 2.13 9 76.93 43569.5 43569.9 20471.6 0.13 0.13 2.13 10 92.31 52298.6 52301.2 24406.5 0.16 0.16 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3722 to 111850 e- Average count rate measured: 566 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.24 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay