*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 09 Mar 2004 13:32:55 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 09 Mar 2004 13:33:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040309 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.01 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040309 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1727.4 1726.9 811.6 0.12 0.13 2.13 2 6.16 3447.6 3447.4 1639.2 -0.16 -0.15 2.10 3 12.31 6901.5 6900.7 3290.9 -0.11 -0.12 2.10 4 18.47 10351.0 10350.4 4978.7 -0.13 -0.14 2.08 5 27.70 15536.3 15534.5 7471.5 -0.07 -0.09 2.08 6 36.93 20724.4 20724.1 10136.9 -0.03 -0.04 2.04 7 46.16 25917.3 25919.8 12524.7 0.01 0.02 2.07 8 61.54 34587.1 34587.8 16679.9 0.10 0.10 2.07 9 76.93 43248.1 43248.6 20375.4 0.13 0.13 2.12 10 92.31 51916.5 51918.0 24283.1 0.16 0.16 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3672 to 110374 e- Average count rate measured: 562 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.17 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay